Brain Slice Chamber:
2, 3, 4, 6 Channel
• 2, 3, 4 and 6 channel slice chambers
• Slices supported on removable insert with nylon net
• Submerged and interface modes by adjusting fluid level
• Proportional Temperature Controller (PTC03) with low noise performance. Available separately.
This slice recording chamber is based on the BSC1 and BSC2
chambers with both "interface" and "submerged"
methods of maintaining slices. Four and six separate chambers
are machined on one acrylic block for simultaneous perfusion
with four or six separate solution lines. As with BSC1 and
BSC2 the temperature is controlled by a proportional heating
unit, the PTC03. Please order this item separately.

The chamber is constructed from clear
acrylic having a diameter of 130mm, overall height is 70mm
and the base plate is 130mm in diameter. The slices rest on
a nylon net held in place by a close fitting removable insert.
Pre-oxygenated medium enters the main body of the chamber
through a fine bore tube which spirals in the heated distilled
water in the lower part of the chamber and enters the upper
part of the chamber directly into the feed for each of the
four or six chambers. Depending on whether submerged
or interface type preparations are required, the height of
the perfusion fluid is adjusted at the exit well by means
of a variable angle stainless steel needle as with our MS-1
chambers.
In the case of interface preparations, the high oxygen tension
is maintained by bubbling a 95% oxygen, 5% carbon
dioxide gas mixture through a sintered bubbler located in
the lower heated part of the chamber. This moistened
and warmed gas mixture enters the upper part of the chamber
via "port holes" and is then deflected by a standard
rectangular glass coverslip across the centrally located
slice preparation. The temperature in the upper chamber is
maintained by ensuring that the medium and moistened gas mixture
enter at the required temperature. This is dependent
on the temperature of the lower chamber body which is warmed
by a heating element controlled by the Proportional Temperature
Controller, PTC03.
|
 |

|  |